Thismenuallows the steppermotor delay time (inmilliseconds) tobechanged. This time is the
duration of pause at the end of onemotormovement and the start of the nextmotormovement. The
factorydefault is 100ms. TheLeft/Right arrowswill change thevalue. PressEnter to exit.
To test the new time settingyoumust scroll toone of themotor axismovementmenu lines. The best
one isprobably theLeft/Right axis, follow the screenprompts to accessmoving "all"motors.
8.3.4 MotorSpeeds
This screenallows the service technicianaccess to settings for all themotor speeds. This information
shouldbeused for referenceonly.
ONLYTRAINEDSERVICETECHNICIANSSHOULD
ATTEMPTTOMAKEMOTORSETTINGCHANGES!
TheMotor speed settings are set and
confirmedwhen the individual axismenu isselected.
In theMotor Speedsmenu, each axis canbemoved for testing, and if necessary, the speedvalues
changed. This allows eachof themovement parameters tobealtered, ramping for beginningand
ending (F; delaybetween steps for starting rate), (R; delaybetween steps for final rate), (S;
accelerationor ramping speed). Once these settings are entered theMotor Speedmenuwill allow the
user the check the settings.
The settings are:
Speed...this is the final speedof themotor.
InitialRate (Initrat)...this is theplunger's initialmovement speed.
Acceleration (Accel)...this is the speed forbeginning (accelerate) andending (decelerate)
motor speeds for the balance of themotors.
!
WARNING FORTRAINED
TECHS ONLY!