SRI GCS Literature Detailed ex SRI website 2014 - page 870

GC ACCESSORIES
Methanizer
Stand-AloneMethanizerAccessory
TheMethanizer ispackedwithanickelcatalyst
powder.Duringanalysis, theMethanizerisheated
to380
o
C. This temperature isset at the factory,
andshouldnotnormallyrequireuseradjustment.
Onceyou turn theMethanizerON, let itwarmup
for at least two minutes before verifying the
temperaturesetpoint.When thecolumneffluent
mixeswiththehydrogencarrierorFIDsupplyand
passes through theMethanizer,COandCO
2
are
converted tomethane. Since theconversionof
COandCO
2
tomethaneoccursafter thesample
compoundshavepassedthroughthecolumn,their
retentiontimesareunchanged. Hydrocarbonspass
throughtheMethanizerunaffected.
TheMethanizeraccessoryenablesanyGCequipped
withaFlameIonizationDetector todetect low levels
of CO and CO
2
. It comes complete with its own
temperaturecontrolboxanduniversalpowersupply,
whichcanoperateonanyofthevariousvoltagesaround
theworld(100-240V).Equippedwithswagelok type
fittings, theMethanizercanbeconnectedanywhere
between thecolumnand theFIDdetector.
Like the FID detector, the Methanizer requires
hydrogen for operation. The required flow rate of
hydrogenthroughtheMethanizeris25mLs/minute. This
canbeacombinationofcarrierandmake-upgases.
Connectyourhydrogensupply to the1/16”make-up
lineon the1/8”bulkheadfitting.Thebulkheadfitting
mayalsobeused tomount theMethanizer bodyon
yourinstrument.
Temperature
display selector switch
HeatON/OFFswitch
Actual
temperature
display
Trimpot for
temperature
setpoint
adjustment
Column
Carrier
Hydrogen
Methanizer
FIDDetector
Bulkhead
fitting
Methanizer
Temperature
control box
Powersupply
Hydrogensupply line
withbulkhead fitting