Operation of theCobraAutosampler on the
8610V gas chromatograph
TheCobra and 8610VGC are now
setup, so the sequence is as fol-
lows:
1) Operator clicksRelayB on
PeakSimple screen.
2)
GC ready light comes on.
This tellsCobra to begin injection
sequence.
3) At themoment theCobra sy-
ringe enters theGC injection
port, theanalysis begins. Peak-
Simple turns onRelayHwhich
inhibits theCobra from begin-
ning another injection.
4)
RelayH turns off just before
analysis ends. GC cools down
and stabilizes. When theGC
ready light turns on again the next
sample is injected. This is re-
peated until all the samples are in-
jected.