SRI GCS Literature Detailed ex SRI website 2014 - page 163

22
GN
71 78 13 13 Local
Resume runningat beginningof current cycle, after
dumping syringe contents towaste, or return to local
control.
AB
65 66 13 13 Local
Abort currentmethod, dump syringe and return to local
control
Q
81 13 13 13 Remote
Echo "ok", useful for synchronizingwithCobra,works
toestablishcommunications initially, anddoesnoharm
ifcommunicationsarealreadyworking
<ESC>
27 13 13 13 Remote
Return cobra to local keypad control, and continue
method if onewas interrupted (Same effect adGO)
For bothRn andGn, theCobrawill transmit "RunComplete"CRwhen themethod is complete. The
systemwillwait for akeypadpress, or any character over the serial link, and then return control to
theCobra. A remote programwill have to re-sendQ toget back to control. All commands should
be terminatedwith aCarriageReturn (binary13). Allmethod#'s (n) shouldbebinary, notASCII
value, i.e. 77824901 forMR1, except theRn command, inwhich then canbe either theASCII
equivalent, or the binaryvalue, i.e. 82011313 and82491313will both start Run1.
10.3 CobraL/SCommandDetails
To program amethod remotely from a host PC:
1) PC sends aQ toget sampler's attention
2) PCwaits for ok, indicatingCobra is ready
3) PC sends "MWn"CR, to indicatemethod is tobe programmed.
4) PCwaits for okCR, indicating theCobra is ready
5)PC sends the followingbinary string, allword (2byte) values:
Byte#
#Bytes
Content
AllowedRange
(PC codemust enforce these limits)
0
2
FirstVial
1-120
2
2
LastVial
1-120
4
2
Numof repeats/sample
1-100
6
2
#of rinseones
0-20
8
2
#of rinse twos
0-20
10
2
rinsevolume
0-100%
12
2
Rinse fill rate
0-100%
14
2
StandardVolume
0-100%
16
2
SampleVolume
0-100%
18
2
Numof solventdumps/inject
0-10
20
2
Numof solv. flushes/inject
0-100