SRI GCS Literature Detailed ex SRI website 2014 - page 624

Installing the supplemental cooling blower on the
OvenFanMotor inSRI 8610C and #10CGCs
SRI 8610CGCs built after Oct
2011 incorporate a cooling blower
which cools the shaft of the oven
fanmotor in order to extend the
operating lifetime. This cooling
blower can be installed on earlier
8610C or 310CGCs.
Remove the 6 screwswhich hold
the bottom cover and tilt theGC on
its back. The 8610CGC is shown
in the photos, but the 310CGC is
substantially similar.
Remove the 4 screws holding the
internal circulation fan bracket.
Un-clip thewhitemolex connector
and remove the bracket to improve
access to the cooling fan location.