Enabling thePressure program on the
SRI 8610CGC
Using a screwdriver adjust the
Carrier Local Setpoint to 000. The
Total Setpoint will now display
whatever Channel Two pressure
program is set to.
TheTotal Setpoint is the sumof
the local setpoint ( the screwdriver
setpoint ) and the remote setpoint
( the setpoint fromPeakSimple ).
If the local setpoint is set to some-
thing other than 000 it will add to
theTotal Setpoint. So if the local
setpoint was 5 and the remote set-
point was 5, theTotal Setpoint
would read 10.
When theGC is first turnedon
andbeforePeakSimple estab-
lishes communication, or if
Channel Two isnot activated the
remote setpoint defaults to
45psi.
Page3