Control Panel Functions
34 of 82
Auto Evacuate SP
(mbar)
The set point that the Antechamber will be evacuated to.
Antechamber Cycles
Determines the number of evacuate/refill cycles that occur under Automatic
operation.
Auto Refill Duration
(sec)
Determines the duration of all refills except the final refill.
Auto Final Refill
Determines the duration of the final refill.
4.3.9 System Config
Pressing the
System Config
button brings up the following screen
:
Figure 24 Glove Box Config screen
Working Gas Type
Allows selection of Nitrogen or Argon as working gas.
Column in Use
Indicates which purifier column is currently in use.
Blower Restart Status
When enabled, the blower will automatically restart after a power failure.
Vacuum Pump Timer
When enabled, the vacuum pump will automatically power down after 10
minutes. The vacuum pump will be automatically powered on as required
by the system.
Working Minimum SP
The lower working pressure set point.
Working Maximum
SP
The higher working pressure set point.
Note:
The PLC control system will maintain the Glovebox pressure
between these Minimum and Maximum set points.
IT Info Page
Contains contact information for customer support.
Screen Setup
Allows the contrast of the HMI screen to be adjusted.