System Operation
42 of 82
4
Close outside door.
5
Turn 3-way valve to evacuate.
6
Evacuate to best level of vacuum (-29 hg).
7
Turn 3-way valve to refill.
8
Repeat steps 5 through 7 at least three times.
9
Close 3-way valve.
10
Open inner door and remove object from chamber.
5.4.2 Removing an Item from the Glovebox Using the Mini Antechamber
1
Ensure that the outside door is fully closed.
2
Ensure that the 3-way valve for mini is closed.
3
Open inner door and insert the object.
4
Close inner door.
5
Open the outer door and remove the object.
Since the chambers share common vacuum and gas piping, we do not recommend using both at
the same time.
5.5 Automatic Antechamber Control
This feature is only displayed if your system has been factory-fitted with automatic
Antechamber control.
Pressing the “Automatic Antechamber Control” button brings up the following screen:
Figure 34 Antechamber Control screen