System Operation
50 of 82
Figure 44 Gas purification module
3
Close the isolation valves (labeled B & C) connected to the solvent column. This is located
in the Gas Purifier. The top must be removed to access valve.
4
From the Main Control Screen select Regen Control. Then select Solvent Regen.
Figure 45 Solvent Regeneration Control screen
5
Press the START button
.
6
Press the
Yes
button next to “Are the Column Valves Closed?” A “click” will be heard as the
solenoid valves open to allow regeneration gas to flow.
7
Adjust needle valve on the regeneration flow meter located on the front of the gas
purification module to set the flow to 25-30 scfh on the graduated scale on the flow meter.