SRI GCS Literature Detailed ex SRI website 2014 - page 95

SRI 8610CGasChromatograph
MultipleGas #3GC configuration
ApplicationExample-2 step process
The same sample is re-
injected and the chroma-
togram shown at right ap-
pears.
Watch the chromatogram
as the peaks appear.
When you see theCO
peak finally elute and the
baseline stabilize ( about
8.5minutes in this exam-
ple ), manually turnRelay
A andRelayG off by
clicking on the lettersA
andG in theRelay/Pump
window. TheRelay/
Pumpwindow can be
made visible or hidden by
clickingAcquisition/
Relay/Pumpwindow.
After RelaysA andG aremanually
turned off, theCO2, Ethylene/Acetylene,
Ethane andWater peaks appear. These
peaks had been stored temporarily in
theHaysep-D column by the action of
theStop-FlowSolenoid.
TheEvent Table ismodified once again
so that RelaysA andG are turned off
automatically at 8.5minutes, so that no
manual actions are required in future
chromatograms.
MultiGas#3
Instructions
Page12
RelaysA&G are
manually turned off
at 8.5minutes
ClickRelayA&G on the
Relay/Pumpwindowwith your
mouse to turn them off