SRI GCS Literature Detailed ex SRI website 2014 - page 89

SRI 8610CGasChromatograph
MultipleGas #3GC configuration
Another example isNatural Gas. Set
theEvent table up to inject and then im-
mediately rotate the valve back to Load
after .1minutes in the Inject position.
This has the effect of performing the
analysis as if Column2was not in the
system. We call this the “ TimingMode”
Column 1 is a 3’ HaysepD andColumn
2 is a 6’ MS13X. TheHaysepD does
not separateOxygen andNitrogen or
CO. Set theStop-Flow solenoid time by
finding the gap betweenMethane and
CO2, in this case about .9minutes.
MultiGas#3
Instructions
Page6
Col 1
Col 1
Col 2
Col 2
Solenoid on at .9
minutes
Solenoid
off
Thegapbetween
Methane andCO2 is a
good time to actuate
theStop-Flow solenoid
TCDdetector
Zoom in