SRI GCS Literature Detailed ex SRI website 2014 - page 102

System Overview
PRE-CONFIGURED GCs
Mud-logger
TheMud-loggerGCsystem isdesigned toprovideacontinous readingof totalhydrocarbons inagas
stream,whileperiodicallyperformingachromatographicseparationofthesampletodeterminethecomposition
of thesamplegasstream.This isaccomplished
witha flow restrictor for themonitoringFID
(#2), and 10-port gas sampling valve&
sample loopfor thespeciationFID(#1).
In the temperature-programmable
columnoven, the 3-meterHayesep-D
packedcolumnefficientlyseparates the
constituenthydrocarbonspeaks(C
1
-C
6
).
In theheatedvalveoven, thesample inlet issplitbetween the10-port
gassamplingvalveandtherestrictor.Therestrictorcontrolsthesample
flowrateforthehydrocarbonmonitoringFID(#2).Thegassampling
valve periodically injects the
contentsofthesampleloopontothe
column for speciationbyFID#1.
In the temperature-programmable
(ambient to400
o
C)columnoven,
the 3-meter Hayesep-D packed
column efficiently separates the
hydrocarbons up toC
6
. Valve injections are automated by the built-in
PeakSimpledatasystem,and the timing isdeterminedby theuser.
Valveoven
Restrictor
Valve
Sample in
andout
Dual FID
detectors
Hayesep-D
packedcolumn
Column
oven
Valve
Restrictor
Sample in Sampleout
30µL sample
loop
3-meterHayesep-D
packedcolumn
FID#1
FID#2
DesignatedasFID#1, thespeciationFIDsignaloccupieschannel1.FID#2, the total
hydrocarbonmonitor, isdisplayedonchannel2.UsingDataloggermode,PeakSimple
willdisplayascaledandcalibratedresult in largenumbersforat-a-glancevisibility.
PeakSimplehasanalarm functionwhichyoucanset toalertyouwhen thesystem
receivesout-of-rangereadings.Abuilt-in,“whisperquiet”aircompressorsuppliesthe
combustionairforbothFIDdetectors.