SRI GCS Literature Detailed ex SRI website 2014 - page 641

Fast Mudlogger GC upgrade kit installation
Oct 2012
TheFast Mudlogger upgrade kit
SRI# 8660-0066 includes thehard-
ware to retrofit anexistingmudlog-
ger for faster analysis times. The
kit includes:
OneSRI tubing cutter
One 1/16” .005 id tubewith split restrictor
15mMXT-1 5micron .53mm capillary column
One 1/16” tubewithValco nut and ferrule
1/8” stainless steel tee fitting
Three nuts andgraphite ferrules
One 35ul loop for thegas sampling valve
Install the 35ul loopbetweenports
3 and10on theValco valve. The
port numbers are stamped into the
valve body next toeachhole.
Put a pieceof tapeover the endof
the tubebefore sliding it through
the ovenwall. The tapewill keep
insulationparticles fromgetting in-
side the tube.