SRI GCS Literature Detailed ex SRI website 2014 - page 643

FastMudlogger GC upgrade kit installation
Oct 2012
Attach the tube to the “tee” fitting.
Connect the column to the tee so
the column slides all theway
through the teeand into the tube
fromport 9or the valve.
This is important because the
samplewill be split and to get a
clean split the columnmust be in-
side the tube.
Connect theother endof the col-
umn to the pre-existing tubegoing
to port 5of the valve.
Connect the split vent tube ( the
tubewith very tiny insidediameter
( .005 inch ) to themiddle of the
tee. Route the split vent outside
the columnoven through ahole in
the right side. Drill a hole if none
exists.
Page3