C02

TRACE WATER ANALYSIS

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argon can be used as well)

• Columns type: 30 meters Plot MTX

Q-

Bond 0.53mmOD

• Carrier flow: 4sccm

• Sampling loop volume: 60uL

• Column temperature: 140 Celcius isothermal

• Impurities concentration: H20: 211 ppb

o

2

4

6

8

10

12

14

16

18

20

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argon can be used as well)

• Columns type: 30 meters Plot MTX

Q-

Bond 0.53mm1D

C4Hs • Carrier flow rate: 4sccm

• Sampling loop volume: 60uL

• Columntemperature: 60 Gelcius isothermal until minute

3, and then 140 Celcius until minute 10.

• Impurities concentration:

C2H2 : 4.1 ppm

C2H4 : 3.1 ppm

C2H6 : 6.1 ppm

CaH6 : 6.0 ppm

CaHs: 5.5 ppm

C4Hs: 5.0 ppm