Background Image
Table of Contents Table of Contents
Previous Page  7 / 47 Next Page
Information
Show Menu
Previous Page 7 / 47 Next Page
Page Background

Water

Acetaldehyde

TRACE ACETALDEHYDE ANALYSIS

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argon can be used as well)

• Columns type: 10 meters Plot Q-Bond 0.32mm1D

• Carrier flow rate: 3sccm

• Sampling loop volume: 60uL

• Column temperature: 80 Celcius isothermal

• Impurities concentration: Acetaldehyde: 132 ppb

o

2

4

6

8

10

12

14

16

18

20

TRACE FORMALDEHYDE

Be

ACETALDEHYDE ANALYSIS

Formaldehyde

Water

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argon can be used as well)

• Columns type: 10 meters Plot Q- Bond 0.32mm1D

• Carrier flow rate: 3sccm

• Sampling loop volume: 60uL

• Column temperature: 70 Celcius isothermal

• Impurities concentration: Formaldehyde: 90 ppb

Acetaldehyde: 2 ppb

o

2

4

6

8

10

12

14

~ROIIII~IJ'1tIC: +61(0)3 9762 2034

Australian Distributors

Id k

~

Importers

&

Manufacurers

www.ete.com

_____

ECHI7~'~g'yPtyLtd

www.chromtech.net.au

Website NEW:

www.chromalytic.com.au

E-mail:

Info@Chromtech.net.au

Tel: 03 9762 2034 ..• in AUSTRALIA