Water
Acetaldehyde
TRACE ACETALDEHYDE ANALYSIS
• Instrument: Micro GCMultiDetek series
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Helium (Argon can be used as well)
• Columns type: 10 meters Plot Q-Bond 0.32mm1D
• Carrier flow rate: 3sccm
• Sampling loop volume: 60uL
• Column temperature: 80 Celcius isothermal
• Impurities concentration: Acetaldehyde: 132 ppb
o
2
4
6
8
10
12
14
16
18
20
TRACE FORMALDEHYDE
Be
ACETALDEHYDE ANALYSIS
Formaldehyde
Water
• Instrument: Micro GCMultiDetek series
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Helium (Argon can be used as well)
• Columns type: 10 meters Plot Q- Bond 0.32mm1D
• Carrier flow rate: 3sccm
• Sampling loop volume: 60uL
• Column temperature: 70 Celcius isothermal
• Impurities concentration: Formaldehyde: 90 ppb
Acetaldehyde: 2 ppb
o
2
4
6
8
10
12
14
~ROIIII~IJ'1tIC: +61(0)3 9762 2034