![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0010.jpg)
TRACE LIGHT HYDROCARBONS ANALYSIS
• Instrument: Micro GCMultiDetek series
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Helium (Argon can be used as well)
• Columns type: 10 meters Plot Q-Bond 0.32mm1D
• Carrier flow rate: 3sccm
• Sampling loop volume: 60uL
• Column temperature: 60 Celcius until min 3. then 130 Celcius
• Impurities concentration:
CH4: 10 ppm
C2H2 : 10 ppm
C2H4 : 10 ppm
C2H6 : 10 ppm
C3H6 : 10 ppm
C3Hs : 10 ppm
C4Hs : 10 ppm
o
4
6
8
10
12
TRACE H2S-COS ANALYSIS IN SYNGAS ORAIR
• Instrument: Micro GCMultiDetek series
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Helium (Argon can be used as well)
• Columns type: 10 meters Plot Q-Bond 0.32mm1D
• Carrier flow rate: 3sccm
• Sampling loop volume: 60uL
• Column temperature: 30 Celcius isothermal
• Impurities concentration:
H2S : 6.4 ppm
COS: 5.1 ppm
H2S
COS
\~
WATER
-
r>.
1"-
I
I
0
2
4
6
8
10
12
14
16
www.ldetek.com• Wher
~
ROIIII
~1J'1tIC:
+61(0)3 97622034
Australian Distributors
~
Importers
&
Manufacurers
_____
ECHI7~'~g'yPtyLtd
www.chrDmtech.net.auWebsite NEW :
www.chromalytic.com.auE-mail :
Info@Chromtech.net.auTel: 03 9762 2034 . . • in AUSTRALIA