![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0014.jpg)
TRACE NHa ANALYSIS
• Instrument: Micro GCMultiDetek series
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Helium (Argoncan be used as well)
• Columns type: 30 meters Plot MTXQ-Bond 0.53mm1D
• Carrier flow rate: 4sccm
• Sampling loop volume: 60uL
• Column temperature: 55 Celcius isothermal
• Impurities concentration:
NH3: 3.4 ppm
20
18
16
14
12
10
2
N~~
~---"---=J/~i'""1 ---~----=------r---.--------,-----,
046 8
TRACE NHa ANALYSIS IN N2 MATRIX
• Instrument: Micro GC MultiDetek series
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Helium (Argon can be used as well)
• Columns type: 30 meters Plot MTXQ-Bond 0.53mm1D
• Carrier flow rate: 4sccm
• Sampling loop volume: 60uL
• Column temperature: 80 Celcius isothermal
• Impurities concentration:
NH3: 11.4 ppm
o
5
10
15
20
25
30
www.ldetek.com• Wher
~