TRACE NHa ANALYSIS

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argoncan be used as well)

• Columns type: 30 meters Plot MTXQ-Bond 0.53mm1D

• Carrier flow rate: 4sccm

• Sampling loop volume: 60uL

• Column temperature: 55 Celcius isothermal

• Impurities concentration:

NH3: 3.4 ppm

20

18

16

14

12

10

2

N~~

~---"---=J/~i'""1 ---~----=------r---.--------,-----,

046 8

TRACE NHa ANALYSIS IN N2 MATRIX

• Instrument: Micro GC MultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argon can be used as well)

• Columns type: 30 meters Plot MTXQ-Bond 0.53mm1D

• Carrier flow rate: 4sccm

• Sampling loop volume: 60uL

• Column temperature: 80 Celcius isothermal

• Impurities concentration:

NH3: 11.4 ppm