AsH3

TRACE AsH3 ANALYSIS

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argon can be used as well)

• Columns type: 30 meter 0.53mm1D Plot Alumina BOND/Na2S04

• Carrier flow rate: 4sccm

• Sampling loop volume: 60uL

• Column temperature: 40 Celcius isothermal

• Impurities concentration:

AsH3: 6.2 ppm

o

2

4

6

8

10

12

14

16

18

20

TRACE PH3 ANALYSIS

• Instrument: Micro GC MultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argoncan be used as well)

• Columns type: 30 meter 0.53mm1D Plot Alumina BOND/Na2S04

• Carrier flow rate: 4sccm

• Sampling loop volume: 60uL

• Column temperature: 40 Celcius isothermal

• Impurities concentration:

PH3: 7.5 ppm

o

2

4

6

8

10

12

14

16

18

20