TRACE AR IN OXYGEN MATRIX

Ar

• Trace Ar in Oxygen matrix

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium

• Column type: 8 feet, 1/8" Packed Stainless Steel ArgoTek

(LDetek's column)

• Carrier flow rate: 20sccm

• Sampling loop volume: 50ul

• Column temperature: 45 Celcius isothermal

• Impurities concentration:

Ar: 0.9 ppm

Oxygen matrix

o

2

4

6

8

10

12

14

16

18

20

TRACE H2S-COS ANALYSIS IN NITROGEN MATRIX

Nitrogen matrix

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Argon (Helium can be used as well)

• Columns type: 1 meter micro packed 1/16"OD Rt-XLSulfurs

• Carrier flow rate: 8sccm

• Sampling loop volume: 200ul

• Column temperature: 60 Celcius isothermal

• Impurities concentration:

H2S: 9.8 ppm

COS: 9.4 ppm