TRACE AR IN OXYGEN MATRIX
Ar
• Trace Ar in Oxygen matrix
• Instrument: Micro GCMultiDetek series
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Helium
• Column type: 8 feet, 1/8" Packed Stainless Steel ArgoTek
(LDetek's column)
• Carrier flow rate: 20sccm
• Sampling loop volume: 50ul
• Column temperature: 45 Celcius isothermal
• Impurities concentration:
Ar: 0.9 ppm
Oxygen matrix
o
2
4
6
8
10
12
14
16
18
20
TRACE H2S-COS ANALYSIS IN NITROGEN MATRIX
Nitrogen matrix
• Instrument: Micro GCMultiDetek series
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Argon (Helium can be used as well)
• Columns type: 1 meter micro packed 1/16"OD Rt-XLSulfurs
• Carrier flow rate: 8sccm
• Sampling loop volume: 200ul
• Column temperature: 60 Celcius isothermal
• Impurities concentration:
H2S: 9.8 ppm
COS: 9.4 ppm
~ROIIII~IJ'1tIC: +61(0)3 9762 2034
Australian Distributors
Id k~
Importers
&
Manufacurers
•
www.ete.com_____
ECHI7~'~g'yPtyLtd
www.chromtech.net.auWebsite NEW:
www.chromalytic.com.auE-mail:
Info@Chromtech.net.auTel: 03 9762 2034 ..• in AUSTRALIA