O
2
matrix
TRACE N2 IN CRUDE ARGON
• Instrument: Micro GCMultiDetek series
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Argon
• Columns type: 2 x 8 feet, 1/8" packed Stainless Steel Molecular Sieve
• Carrier flow rate: 30sccm
• Sampling loop volume: 200uL
• Column temperature: 60 Celcius isothermal
• Impurities concentration:
N2 : 98 ppm
o
2
4
6
8
10
12
14
16
18
20
• Instrument: Micro GCMultiDetek series
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Helium (Argoncan be used as well)
• Columns type: 2 X 10', 1/8" PackedStainless Steel Hayesep D
• Carrier flow rate: 30sccm
N
20
• Sampling loop volume: 1mL
• Column temperature: 50 Celcius isothermal
• Impurities concentration:
CH4: 8.1 ppm
C02: 50.7 ppm
N20 : 10.0 ppm
o
5
10
15
20
25