TRACE FORMALDEHYDE, ACETALDEHYDE ANALYSIS

Formaldehyde

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argoncan be used as well)

• Columns type: 30 meters Plot MTXQ-Bond 0.53mm1D

• Carrier flow rate: 4sccm

• Sampling loop volume: 60uL

• Column temperature: 100 Celcius isothermal

• Impurities concentration:

Formaldehyde: 8.1 ppb

Acetaldehyde: 0.3 ppb

r

T

o

2

4

6

8

10

12

14

16

18

20

TRACE ACETALDEHYDE ANALYSIS

Acetaldehyde

• Instrument: Micro GC MultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argon can be used as well)

• Columns type: 30 meters Plot MTXQ-Bond 0.53mm1D

• Carrier flow rate: 4sccm

• Sampling loop volume: 60uL

• Column temperature: 100 Celcius isothermal

• Impurities concentration:

Acetaldehyde: 132ppb