TRACE H2S-COS ANALYSIS IN SYNGAS ORAIR

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argon can be used as well)

• Columns type: 30 meters Plot MTX-Q-Bond 0.53mm1D

• Carrier flow rate: 4sccm

• Sampling loop volume: 60uL

• Column temperature: 100 Celcius isothermal

• Impurities concentration:

H2S : 6.4 ppm

COS: 5.1 ppm

H2S

WATER

t

s

A

2

4

6

8

10

12

14

TRACE CH4-C02 analysis

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argon can be used as well)

• Columns type: 30 meters Plot MTXQ-Bond0.53mm1D

• Carrier flow rate: 4sccm

• Sampling loop volume: 60uL

• Column temperature: 30 Celcius isothermal

• Impurities concentration:

CH4 : 9.0 ppm

C02: 3.1 ppm