Background Image
Table of Contents Table of Contents
Previous Page  16 / 47 Next Page
Information
Show Menu
Previous Page 16 / 47 Next Page
Page Background

TRACE N20 ANALYSIS IN ARGON MATRIX

• Instrument: Micro GCMultiDetek series

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argon can be used as well)

• Columns type: 30 meter 0.53mm1D Plot Alumina BONDjNa2S04

• Carrier flow rate: 4sccm

• Sampling loop volume: 60ul

• Column temperature: 40 Celcius isothermal

• Impurities concentration:

N20: 5.4 ppm

o

2

4

6

8

10

12

14

16

18

20

TRACE C2'S-C3'S-C4'S ANALYSIS IN N2 MATRIX

• Instrument: Micro GC MultiDetek series

C3HS

• Detector: PlasmaDetek series (PED)

• Carrier Gas: Helium (Argon can be used as well)

C2Ha

C3Ha

• Columns type: 30 meter 0.53mm1D Plot Alumina BONDjNa2S04

• Carrier flow rate: 4sccm

C2H4

C4HS • Sampling loop volume: 60ul

• Column temperature: 40 Celcius isothermal until minute 3, and

then 150 untill minute 8

• Impurities concentration:

C2Ha: 5.1 ppm

C2H4 : 4.0 ppm

C3HS : 6.6 ppm

C3Ha : 5.9 ppm

C2H2 : 4.0 ppm

C4HS : 6.5 ppm

o

10

12

14

16

18

20

www.ldetek.com