![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0016.jpg)
TRACE N20 ANALYSIS IN ARGON MATRIX
• Instrument: Micro GCMultiDetek series
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Helium (Argon can be used as well)
• Columns type: 30 meter 0.53mm1D Plot Alumina BONDjNa2S04
• Carrier flow rate: 4sccm
• Sampling loop volume: 60ul
• Column temperature: 40 Celcius isothermal
• Impurities concentration:
N20: 5.4 ppm
o
2
4
6
8
10
12
14
16
18
20
TRACE C2'S-C3'S-C4'S ANALYSIS IN N2 MATRIX
• Instrument: Micro GC MultiDetek series
C3HS
• Detector: PlasmaDetek series (PED)
• Carrier Gas: Helium (Argon can be used as well)
C2Ha
C3Ha
• Columns type: 30 meter 0.53mm1D Plot Alumina BONDjNa2S04
• Carrier flow rate: 4sccm
C2H4
C4HS • Sampling loop volume: 60ul
• Column temperature: 40 Celcius isothermal until minute 3, and
then 150 untill minute 8
• Impurities concentration:
C2Ha: 5.1 ppm
C2H4 : 4.0 ppm
C3HS : 6.6 ppm
C3Ha : 5.9 ppm
C2H2 : 4.0 ppm
C4HS : 6.5 ppm
o
10
12
14
16
18
20
www.ldetek.com• Wher
~
ROIIII
~1J'1tIC:
+61(0)3 97622034
Australian Distributors
~
Importers
&
Manufacurers
_____
ECHI7~'~g'yPtyLtd
www.chrDmtech.net.auWebsite NEW:
www.chromalytic.com.auE-mail:
Info@Chromtech.net.auTel: 03 9762 2034 ..• in AUSTRALIA