EPAMethod 16BGC configuration
November 2013
Inject a sample containing oth-
er sulfur species. Herewe
have injectedH2S, COS and
DMS but noSO2. Identify and
calibrate the peaks if neces-
sary. Herewe show 40ppm
eachH2S, COS andDMS ( di-
methylsulfide ).
Plug in the reactor.
Then verify that the reactor tempera-
ture increases to about 260. Thiswill
take several minutes. The real tem-
perature of the reactor is about
1000C, as the readout temperature
must bemultiplied by a factor of four.