SRI GCS Literature Detailed ex SRI website 2014 - page 493

Do the same thing for theTotal Diglycerides
andTotal Triglycerides peaks. Highlight the
peak, click theCalibrate button and then load
TheDiOlein calibration curve for theTotal
Diglycerideswindow.
Do the same for theTotal Triglycerideswin-
dow, except load theTriOlein calibration
curve.
TheComponent table should now look like
the one at right.
Double check that theTotal Mono, Total Di
andTotal Tri component windows have the
correctmultiplication factors inserted.
Be sure to click File/SaveAll to ensure that
all thework you have done so far is saved.
PerformingASTM6584 freeand total glycerin in
BioDiesel usinganSRI GasChromatograph
andPeakSimple software version3.67
Page 12