SRI GCS Literature Detailed ex SRI website 2014 - page 488

Navigate to theComponent
screen for Channel 1.
TheComponent screen
should look like the one to the
right except that the names of
the calibration curves for each
component will not yet have
been entered.
The internal standard
peaks ( butanetriol and tri-
caprin ) do not get calibra-
tion curves.
Notice that the peak num-
ber is followed by a number
in parentheses.
The number in parentheses
indicateswhich peak is the
internal standard. So
Monolienwhich is peak#3
uses tricaprin ( peak#4 ) as
its internal standard. Glyc-
erol which is peak#1 uses
butanetriol ( peak#2 ) as its
internal standard.
Save theComponent ta-
ble so youdon’t have to
enter this information
over again.
PerformingASTM6584 freeand total glycerin in
BioDiesel usinganSRI GasChromatograph
andPeakSimple software version3.67
Page 7