SRI GCS Literature Detailed ex SRI website 2014 - page 492

TheComponent Table should now look like
the one shown at right. Note that only the
four standard peaks ( glycerin, Mono, di and
Tri ) have calibration curves attached.
The two internal standards ( butanetriol and
tricaprin ) do not have calibration curves.
Also the 3 total peaks ( Total Monoglyc-
erides, Total Diglycerides andTotal Triglycer-
dies do not as yet have calibration curves at-
tached to the componentswindows.
Highlight theTotal Monoglycerides compo-
nent by clicking it with yourmouse,
Then click theCalibrate button on the lower
right of the screen.
A blank calibration curvewindowwill appear.
Click the Load button and then select the
same calibration curvewhich you created for
MonoOlein.
When you return to theComponent table you
will see that theTotal Monoglycerides com-
ponenet now has the same calibration curve
attached as theMonoOlein component.
PerformingASTM6584 freeand total glycerin in
BioDiesel usinganSRI GasChromatograph
andPeakSimple software version3.67
Page 11