SRI GCS Literature Detailed ex SRI website 2014 - page 479

Onceall thepeaks are cali-
brated, the component table
should look like theone to the
right. Notice that theTotal
Mono, Total Di, andTotal Tri
components use the same
calibration curves as the indi-
vidual Mono, Di andTriolien
peaks. Thismakes sensebe-
cause these compounds are
chosen in theASTM6584
method tobe representativeof
the rangeof glyceride com-
poundswhichmay be found in
actual samples.
Navigate to theResults
screen and verify that the
results look like the screen to
the right. ( for a level 5 cali-
bration standard ) Notice that
the internal standard result
for the Total Mono, Di and
Tri is adjusted by the factor
( ,2591, .1488and .1044 re-
spectively )
PerformingASTM 6584 free and total glycerin in
BioDiesel using anSRI GasChromatograph
andPeakSimple software
1...,469,470,471,472,473,474,475,476,477,478 480,481,482,483,484,485,486,487,488,489,...870
Powered by FlippingBook