6 Maintenance & Servicing
6.1 Impact Device Maintenance
After the impact device has been used for 1000--2000 times, please use the nylon
brush provided to clean the guide tube and impact body. When cleaning the guide tube,
unscrew the support ring first, then take out the impact body, spiral the nylon brush in
counter-clock direction into the bottom of guide tube and take it out for 5 times, and
then install the impact body and support ring again.
Release the impact body after use.
Any lubricant is absolutely prohibited inside the impact device.
6.2 Instrument Maintenance Program
When using standard Rockwell hardness block to testing, if all the error is bigger than 2
HRC, it may be the invalidation of impacted ball top caused by abrasion. Changing the
spherical test tip or impact object should be considered.
When the hardness tester appears some other abnormal phenomena, please do not
dismantle or adjust any fixedly assembled parts. Fill in and present the warranty card to
us. The warranty service can be carried on.
6.3 Fault Analysis & Evacuation
Fault Appearance
Fault Analysis
Handling method
Failure power on
Battery exhaustion
Replace the batteries
No measured value
mpact device cable failure
Replace the cable
6.4 Notice of Transport and Storage Conditions
Keep it away from vibration, strong magnetic field, corrosive medium, dumpiness and
dust. Storage in ordinary temperature.
With original packing, transport is allowed on the third grade highway.