93

Simulated Distillation (C5-C44) on Rtx®-2887

GC COLUMNS |

APPLICATION-SPECIFIC COLUMNS

Petroleum & Petrochemical

Simulated Distillation Analysis (C5-C44)

GC_PC00750

Column

Rtx®-2887, 10 m, 0.53 mm ID, 2.65 µm (cat.# 10199)

Sample

C5 to C44 hydrocarbon standard

Diluent:

Carbon disulfide

Conc.:

0.01-0.1 wt. %

Injection

Inj. Vol.:

1 µL direct

Inj. Temp.:

360 °C

Oven

Oven Temp.:

35 °C to 360 °C at 15 °C/min (hold 5 min)

Carrier Gas

He, constant flow

Flow Rate:

15 mL/min

Linear Velocity:

112 cm/sec

Detector

FID@ 360 °C

Peaks

1. C5

2. C6

3. C7

4. C8

5. C9

6. C10

7. C11

8. C12

9. C14

10. C16

11. C18

12. C20

13. C24

14. C28

15. C32

16. C36

17. C40

18. C44

also available

MXT®-1HT SimDist

and more simulated

distillation products

See

pages 113–115

.

free literature

Rtx®-2887/MXT®-2887

Restek

s Capillary GC Columns for Simulated

Distillation of Petroleum Fractions

Download your free copy from

lit. cat.# 59567B

www.restek.com

Rtx®-2887 Column

(fused silica)

(nonpolar phase; Crossbond® 100% dimethyl polysiloxane—for simulated distillation)

• Application-specific column for simulated distillation.

• Stable to 360 °C.

The Rtx®-2887 column's stationary phase, column dimensions, and film thickness have

been optimized to exceed the resolution and skewing factor requirements specified in

ASTM Method D2887. Each column is individually tested to guarantee a stable base-

line with low bleed and reproducible retention times. The Crossbond® methyl silicone

stationary phase has increased stability compared to packed columns, ensuring stable

baselines and shorter conditioning times.

similar phases

DB-2887, Petrocol 2887, Petrocol EX2887

10-Meter

ID

df

temp. limits

cat.#

price

0.53 mm

2.65 µm

-60 to 360 °C

10199