Table of Contents Table of Contents
Previous Page  90 / 628 Next Page
Information
Show Menu
Previous Page 90 / 628 Next Page
Page Background

88

www.restek.com

GC COLUMNS |

APPLICATION-SPECIFIC COLUMNS

Foods, Flavors, & Fragrances

Rxi®-624Sil MS Columns

(fused silica)

(midpolarity Crossbond® phase)

• Low-bleed, high-thermal stability column—maximum temperatures up to 320 °C.

• Inert—excellent peak shape for a wide range of compounds.

• Selective—G43 phase highly selective for volatile organics and residual solvents,

great choice for USP<467>.

• Manufactured for column-to-column reproducibility—well-suited for validated

methods.

Rxi®-624Sil MS (G43) Structure

x

Si

O

(CH

2

)

3

CH

3

CH

3

CH

3

Si

O

R

O

CH

3

CH

3

Si

94%

y

CH

3

CH

3

Si

C N

similar phases

DB-624, VF-624ms, CP-Select 624 CB

20-Meter

30-Meter

60-Meter

75-Meter

105-Meter

ID df

temp. limits cat.#

price

cat.#

price

cat.#

price

cat.#

price

cat.#

price

0.18 mm 1.00 µm -20 to 300/320 °C 13865

0.25 mm 1.40 µm -20 to 300/320 °C

13868

13869

0.32 mm 1.80 µm -20 to 300/320 °C

13870

13872

0.53 mm 3.00 µm -20 to 280/300 °C

13871

13873

13874

13875

Terpenes Analysis for Cannabis and Hops

Similar to: (6%-cyanopropylphenyl)-methylpolysiloxane

1

2

3

4 5

6

7

8

9

10

11

12

13 14

8

8.2

8.4 8.6

8.8

9

15

16 17

10.6 10.8 11 11.2 11.4 11.6 11.8 12 12.2

18

19 20 21

22 23

24,25 26

27

28

29 30

31

32

33

34

35

36

8

9

10

11

12

13

14

15

16

Time (min)

Medical Cannabis Terpenes on Rxi®-624Sil MS by FET-HS-GC

GC_FS0518

Column

Rxi® -624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)

Sample

Terpenes mix

Diluent:

Isopropyl alcohol

Conc.:

200 ng/µL (0.02% wt/vol). The sample was prepared by placing

10 µL into the headspace vial.

Injection

headspace-loop split (split ratio 10:1)

Liner:

Sky® 1.0 mm ID straight inlet liner (cat.# 23333.1)

Headspace-Loop

Inj. Port Temp.:

250 °C

Instrument:

Tekmar HT-3

Inj. Time:

1.0 min

Transfer Line Temp.: 160 °C

Valve Oven Temp.: 160 °C

Needle Temp.:

140 °C

Sample Temp.:

140 °C

Sample Equil. Time: 30.0 min

Vial Pressure:

20 psi

Loop Pressure:

15 psi

Oven

Oven Temp.:

60 °C (hold 0.10 min) to 300 °C at 12.50 °C/min (hold 3.0 min)

Carrier Gas

He, constant flow

Linear Velocity:

33 cm/sec

Detector

FID@ 320 °C

Make-up Gas

Flow Rate:

45 mL/min

Make-up Gas Type: N

2

Hydrogen flow:

40 mL/min

Air flow:

450 mL/min

Data Rate:

20 Hz

Instrument

Agilent/HP6890 GC

Peaks

t

R

(min)

1.

α

-Pinene

7.39

2. Camphene

7.71

3.

β

-Myrcene

7.98

4. Sabinene

8.02

5.

β

-Pinene

8.11

6.

α

-Phellandrene

8.4

7.

δ

3-Carene

8.44

8.

α

-Terpinene

8.57

9. Ocimene

8.61

10. Limonene

8.71

11.

p

-Cymene

8.75

12.

β

-Ocimene

8.82

13. Eucalyptol

8.91

14.

γ

-Terpinene

9.06

15. Terpinolene

9.47

16. Linalool

9.87

17. Fenchone

10.06

18. Isopulegol

10.73

19. dl-Menthol

11.08

20. Borneol

11.19

21.

α

-Terpineol

11.29

22. Dihydrocarveol

11.40

23. Citronellol

11.51

24. Geraniol

11.82

25. 2-Piperidinone

11.88

26. Citral 1

11.92

27. Pulegone

11.97

28. Citral 2

12.24

29. Citral 3

13.19

30. Citral 4

13.43

31.

β

-caryophyllene

13.83

32.

α

-Humulene

14.21

33. Nerolidol 1

14.78

34. Nerolidol 2

15.08

35. Caryophyllene oxide 15.92

36.

α

-Bisabolol

16.43