92

www.restek.com

GC COLUMNS |

APPLICATION-SPECIFIC COLUMNS

Petroleum & Petrochemical

Detailed Hydrocarbons Analysis on Rtx®-DHA-100

Rtx®-5 DHA Tuning Column

(fused silica)

(Crossbond® 5% diphenyl/95% dimethyl polysiloxane—optimized for hydrocarbon analysis)

Detailed Hydrocarbon Analysis (DHA)

did you know?

Using hydrogen instead of helium can cut

analysis time in half! Visit

www.restek.com

to learn more.

ASTM

Method

D6729

D6730

D6733

D5501

Dimensions

100 m x 0.25 mm, 0.50 µm

100 m x 0.25 mm,

0.50 µmw/ precolumn

50 m x 0.20 mm, 0.50 µm

150 m x 0.25 mm, 1.0 µm

Column

Rtx-DHA-100

Rtx-DHA-100 &

Rtx-5 DHA

Tuning Column

Rtx-DHA-50

Rtx-DHA-150

cat. #

10148

10148

&

10165

10147

10149

Method Recommended

10

20

30

40

50

60

70

80

90

100

Time (min)

1

2

3

4

5

6

7

8

9

11

10

12

13

14 15

16

17

18

19

20

21

22

23

24

25

26

27

28

29 30

GC_PC00743

Column

Rtx®-DHA-100, 100 m, 0.25 mm ID, 0.50 µm (cat.# 10148)

using Rtx®-5 DHA tuning column, 2.62 m, 0.25 mm ID, 1.0 μm (cat.# 10165)

Sample

Custom detailed hydrocarbons analysis (DHA) mix

Conc.:

Neat

Injection

Inj. Vol.:

0.01 µL split (split ratio 150:1)

Liner:

Cup splitter with Siltek® deactivation (cat.# 20709-214.1)

Inj. Temp.:

200 °C

Oven

Oven Temp.:

5 °C (hold 15 min) to 50 °C at 5 °C/min (hold 50 min) to 200 °C at 8 °C/min

(hold 10 min)

Carrier Gas

He, constant flow

Flow Rate:

2.3 mL/min

Linear Velocity:

28 cm/sec

Detector

FID@ 250 °C

Peaks

1. Ethanol

2. C5

3.

tert

-Butanol

4. 2-Methylbutene-2

5. 2,3-Dimethylbutane

6. Methyl

tert

-butyl ether (MTBE)

7. C6

8. 1-Methylcyclopentene

9. Benzene

10. Cyclohexane

11. 3-Ethylpentane

12. 1-

tert

-2-Dimethylcyclopentane

13. C7

14. 2,2,3-Trimethylpentane

15. 2,3,3-Trimethylpentane

16. Toluene

17. C8

18. Ethylbenzene

19.

p

-Xylene

20. 2,3-Dimethylheptane

21. C9

22. 5-Methylnonane

23. 1,2-Methylethylbenzene

24. C10

25. C11 (undecane)

26. 1,2,3,5-Tetramethylbenzene

27. Naphthalene

28. C12 (dodecane)

29. 1-Methylnaphthalene

30. C13 (tridecane)

Resolve all critical pairs—including

2,3,3-trimethylpentane and toluene!

free literature

Detailed Hydrocarbon Analysis

Featuring Rtx®-DHA Columns

Download your

free copy from

lit. cat.#

PCFL1007B-UNV

www.restek.com

Rtx®-DHA Columns

(fused silica)

(Crossbond® 100% dimethyl polysiloxane—optimized for hydrocarbon analysis)

• Columns meet or exceed all ASTM D6730-01 and CAN/CGSB 3.0 No. 14.3-99

method guidelines; test report for method D6730 supplied with each column.

• Excellent responses and peak symmetry for polar oxygenates.

• Stable to 340 °C.

Gasolines are complex mixtures of hundreds of compounds. Information about con-

centrations of the individual components is important for evaluating raw materials

and for controlling refinery processes. ASTM D6730-01 outlines a high-resolution GC

method for detailed hydrocarbon analysis (DHA) of gasolines. Rtx®-DHA columns

are ideal for DHA methods and easily meet or exceed both ASTM D6730-01 and Ca-

nadian General Standards Board CAN/CGSB 3.0 No. 14.3-99 requirements. Every

Rtx®-DHA column is tested for retention, efficiency, stationary phase selectivity, and

bleed—guaranteeing reproducible column-to-column performance.

similar phases

HP-PONA, DB-Petro, CP-Sil PONA C8,

Petrocol DH

50-Meter

100-Meter

150-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

cat.#

price

0.20 mm 0.50 µm -60 to 300/340 °C 10147

0.25 mm 0.50 µm -60 to 300/340 °C

10148

1.00 µm -60 to 280/340 °C

10149

5-Meter

ID

df

temp. limits

cat.#

price

0.25 mm 1.00 µm -60 to 325/350 °C 10165