restekapp07 - page 249

RestekCorporation • (800)356-1688 • (814) 353-1300 •
food& flavor
Figure2
Obtain fine structure resolution by degree of unsaturationwith common oils on a 15-meter Rtx
®
-65TG column.
discrimination and thermal degradation, which is commonwith
normal split/splitless injectors. The column shows excellent
separation and peak symmetry of the triglycerides by both
degree of unsaturation and carbon number. Fine structure
resolution of T50 (PPS/PPO), T52 (PSS/PSO/POO), andT54
(SSS/SSO/SOO/OOO) triglycerides is obtainedwithminimal
column bleed at 365°C. The optimized analysis time is less than
35minutes. Figure 2 shows the analysis of canola, corn, and
olive oils on a 15m, 0.25mm ID, 0.10µmRtx
®
-65TG column.
The unsaturates are all well separated in less than 8minuteswith
minimal column bleed at 365°C.
TheRtx
®
-65TG column is ideal for triglyceride analysis.
Separation by degree of unsaturation aswell as carbon number
is obtained in under 35minuteswithminimal column bleed at
15m, 0.25mm ID, 0.10µmRtx
®
-65TG (cat.# 17005)
0.8µL split injection of triglycerides.
4
6
8
7
9
14
15
12
13
min. 4
8
min. 4
8
8
9
12
13
14
15
min. 4
8
6
4
7
8
12
13
14
1. PPP
2. PPS
3. PPO
4. POP
5. PSO
6. PLP
7. POO
8. PLO
9. PLL
10. SSO
11. OOS
12. OOO
13. OLO
14. OOL
15. LLL
Oven temp.:
a) 355°C (hold 1min.) to 365°C@ 1°C/min.
b) 350°C (hold 0.5min.) to 365°C@ 1°C/min.
c) 350°C (hold 1min.) to 365°C@ 0.5°C/min.
a) canola oil
b) corn oil
c) olive oil
Inj. &det. temp.:
365°C
Carrier gas:
hydrogen
Linear velocity:
50cm/sec.
FID sensitivity:
8 x 10
-11
AFS
Split ratio:
40:1
References:
1. Geeraert and Sandra, “CapillaryGC of Triglycerides in Fats andOils
Using aHighTemperature PhenylMethyl Silicone Stationary Phase,”
Journal of HRC&CC, Vol. 8, Aug 1985, pp. 415-419.
2. Association ofAnalytical Chemists, Official Methods ofAnalysis of the
AOAC, 17th ed.,Method 986.19.
3. AmericanOil Chemists Society, Official Methods and Recommended
Practices (1994), Method Ce 5-86.
365°C. BecauseRtx
®
-65 columns are individually testedwith a
temperature-programmed triglyceride testmixture, lowbleed
and high efficiency are guaranteed. Rtx
®
-65TG columns are
available in both 15- and 30-meter lengths in 0.25, 0.32, and
0.53mm IDswith a 0.10µm film thickness.
1...,239,240,241,242,243,244,245,246,247,248 250,251,252,253,254,255,256,257,258,259,...324
Powered by FlippingBook