ASTMMethodD-6352 is a gas chromatography (GC) method
developed for the determination of petroleum distillates with a
boiling point range of 174
°
C to 700
°
C. Often referred to as
"high-temperature simulated distillation," thismethod requires
a capillary column capable of withstandingGC oven tempera-
tures up to 430
°
C. This presentsmany challenges for analysts
becausemost capillary columns aremanufactured using poly-
imide-coated fused silica tubing.At temperatures above 380
°
C,
even the best polyimide coating becomes brittle, which leads
to very short column lifetimes. In addition, themethyl silicone
stationary phase recommended in themethod alsomust sur-
vive these high temperatures.
TheMXT
®
-1HTSimDist column is amajor improvement in
column technology for high-temperature simulated distillation.
By combining a new, proprietary polymer synthesis technolo-
gy, Siltek
deactivation, and rugged Silcosteel
®
tubing, we
developed a capillary column that meets all the criteria of
Applications
note
Fast, High-Temperature SimDist AnalysesUsing
MXT
®
-1HTSimDist CapillaryGCColumn and theGCRacer
TheMXT-1HTSimDist column offers lower bleed and improvedpeak symmetry compared to other
high temperature columns.
RunConditions: C44-C100Hydrocarbons
Column:
MXT-1HTSimDist
Serial #:
339749B
Description:
5m, 0.53mmID, 0.10µm
Sample:
Polywax
®
1000
Part #:
36227
Solvent:
CarbonDisulfide
SampleSize:
0.2uL
Instrument:
HP6890GC
Injector:
on-column (track oven)
Carrier Gas:
Helium(constant pressure)
HeadPressure:
1.0psi
Linear Velocity:
60cm/sec.
FlowRate:
7.8mL/min.
Detector:
FID/430°C
Make upGas Flow: 40cc/min.
Temp. Program:
40°C to 430°C@10°C/min.
(hold 30min.)
ASTMMethodD-6352. ThisMXT
®
-1HTSimDist column is
available as in a 5m, 0.53mm ID, 0.1µm film dimension to
conform to the requirements of thismethod. It exceeds the res-
olution, peak shape, and bleed criteria for hydrocarbons rang-
ing up toC-110. Because theMXT
®
-1HTSimDist column is
coatedwith a 100% dimethyl polysiloxane polymer, it will
give the correct retention time/boiling point curve. TheMXT
®
-
1HTSimDist column exhibits low bleed and excellent inert-
ness, and the rugged Silcosteel
®
tubingwill hold up to temper-
atures in excess of 430°C.
To demonstrate the lower bleed and improved peak shape of
this innovative column, a Polywax
®
1000 referencematerial
was analyzed using aMXT
®
-1HTSimDist column and two
other columns that are commonly used for this application
(Figure 1)
. TheMXT
®
-1HTSimDist column exhibits lower
bleed and improved peak symmetry compared to competitor
columns.
Figure 1
RestekMXT-1HT Column
Competitor Column A
Competitor Column B
85pA
78pA
28pA
GC_PC00541
GC_PC00542
min. 0
10
20
30
40
50
60
min. 0
10
20
30
40
50
60
min. 0
10
20
30
40
50
60
GC_PC00530
C20
C30
C40
C50
C60
C70
C80
C90
C100
C110
C30
C40
C50
C60
C70
C80 C90
C100
C110
C30
C40
C50
C60
C70
C80
C90
C100
1...,131,132,133,134,135,136,137,138,139,140 142,143,144,145,146,147,148,149,150,151,...324
Powered by FlippingBook