restekapp07 - page 136

1
2
3
4
5
6
7
8
9
10
11
12
13
0
2
4
6
8
10
12
14
14
Figure 2
For permission to reproduce any portion of this application note, please contact Restek’s publications/graphics department by phone (ext. 2128) or FAX.
RestekTrademarks:
Pinnacle, Pinnacle II, and theRestek logo.
andmethanol, the process of switching from theC18 to the
Cyano column is only amatter of removing the primary column
and installing the confirmation column on the sameHPLC sys-
tem. There is relatively little downtime, only that required for
the column to equilibrate.
Notice that all of the coeluting pairs form theC18 column are
resolved from one another on theCyano column. There is a
cluster of compounds: 2-nitrotoluene, 3-nitrotoluene, 4-nitro-
toluene, 1,3,5-trinitrotoluene, and 1,3 dinitrotoluene on the
Cyano column, but these compounds arewell resolved on the
C18 column.Again, the selectivity between the original
The Pinnacle II
Cyano column shows slightly better resolution for explosives compounds and is an excellent confirmation
column to the Pinnacle II
C18 for this analysis.
Pinnacle
Cyano and the newPinnacle II
Cyano columns is
similar, but the Pinnacle II
column shows slightly better reso-
lution.
Restek controls the rawmaterial quality from the very begin-
ning of the silicamanufacturing process.Add our phase bond-
ing and column packing experience to this high level of quality
control, and you benefit from even better column-to-column
and analysis-to-analysis reproducibility. Because of this and
their economical production, Pinnacle II
HPLC columns pro-
vide a cost-effective analytical tool formany traditional meth-
ods used in the environmental industry.
Pinnacle II
C18 5µmColumns
Length
1.0mm ID 2.1mm ID 3.2mm ID 4.6mm ID
50mm
9214551
9214552
9214553
9214555
100mm
9214511
9214512
9214513
9214515
150mm
9214561
9214562
9214563
9214565
250mm
9214571
9214572
9214573
9214575
Conditions:
Mobile phase:
water:methanol (50:50, v/v)
Flow:
1.5mL/min
Temp.:
27.0°C
Det.:
UV@ 254nm
Sample:
Inj.:
3µL
Conc.:
500 µg/mL
each component
Solvent:
acetonitrile
Restek 8330Calibrationmix #1: Catalog # 31450
Restek 8330CalibrationMix #2 : Catalog# 31451
LC_0188
PeakList:
1. nitrobenzene
2. 2-nitrotoluene
3. 4-nitrotoluene
4. 3-nitrotoluene
5. 1,3-dinitrobenzene
6. 1,3,5-trinitrobenzene
7. 2,6 dinitrotoluene
8. 2,4-dinitrotoluene
9. 2,4,6-trinitrotoluene
10. 4-amino-2,6-dinitrotoluene
11. 2-amino-4,6-dinitrotoluene
12. RDX
13. tetryl
14.HMX
Pinnacle II
Cyano 5µmColumns
Length
1.0mm ID 2.1mm ID 3.2mm ID 4.6mm ID
50mm
9216551
9216552
9216553
9216555
100mm
9216511
9216512
9216513
9216515
150mm
9216561
9216562
9216563
9216565
250mm
9216571
9216572
9216573
9216575
LC_0193
Column:
Pinnacle
CN
Catalog #:
9116575
Dimensions:
250 x 4.6mm
Particle size: 5µm
Pore size:
120Å
Column:
Pinnacle IICN
Catalog #:
9216575
Dimensions:
250 x 4.6mm
Particle Size:
5µm
Pore Size:
110Å