SRITN12 - page 40

HPLC
Model 210D
11. UVDetector
TheUVdetectorsignal isconnected tochannel1.Thedetectoruses
alowvolumeflow-throughcell.Thecelltemperatureisfactorysetto
40
o
C,which issuitableforawiderangeofcommonapplications.
Bothdetectorsarehoused in the
cell in themiddleof thisboard.
Note for thosewhohave the4or6channeldatasystemupgrade:
Unlessyouhaveconnectedanexternaldetector to the210D, the
column temperaturewilldisplayonchannel3,and thedetectorcell
temperaturewilldisplayonchannel4.
12. Setting aGradient Program
Clickon theGradientbutton forchannel1. Ifyour system is isocratic, the
Inital
and
Final
percentagesare
100%.Theonlyvariable is the
Hold
timeperiod,which isdeterminedby the lengthofyouranalytical run.
Settingagradientprogramautomaticallystarts thepumpatavery lowspeed.Forbinarygradientsystems,
loadagradientprogramfromyourPeakSimpleprogramfolder(version3.19andlater)byclickingonthe
Load
button in the
Channel1gradientcontrol
window.Use the
Add
button tocreateanewgradient.
Isocraticgradient
Binarygradient