PreconfiguredGCSystems
MultipleGasAnalyzer #3GCSystem
Another example is Natural Gas. Set the Event table up to inject and
then immediately rotate the valve back to Load after .1 minutes in
the Inject position. This has the effect of performing the analysis as
if Column 2 was not in the system. Column 1 is a 3’ HaySep D and
Column 2 is a 6’ MS13X. The HaySep D does not separate Oxygen
and Nitrogen or CO. Set the Stop-Flow solenoid time by inding the
gap betweenMethane and CO
2
, in this case about .9 minutes. With
theEvent tablemodiied, theOxygen,NitrogenandMethaneseparate
on theMS13X. Then theStop-Flowsolenoid isde-energizedandvalve
rotatedback toLoadposition (bothat 4.00minutes) and the remaining
peaks (Ethane, Propane, Water, Butanes, and Pentanes) which were
immobilizedon theHaySepD (column 2) elute normally.
8610-0370
MultipleGasAnalyzer #3GCwithTCDdetector and1channel
PeakSimpledata system
8610-0371
MultipleGasAnalyzer #1withTCD,Methanizer,
FID, built-in “whisper quiet” air compressor and
6channel PeakSimpledata system
8610-0372
MultipleGasAnalyzer #1withTCDandHIDdetectors,
and6channel PeakSimpledata system
OPTIONS&UPGRADES: 6 channel USBPeakSimpledata system, additional gas sampling valve,
additional detectors
(VOLTAGE: for 115VAC, use “part number-1” [ex: 8610-0070-1] for 230VAC, use “part number-2”)