MRC-2013-Catalog - page 237

237
E
ENVIRONMENTAL CHAMBERS
The walk-in environmental test room is composed of the
thermal wallboard by means of pu foaming. It is easy to
disassemble & transport. According to the environmental
conditions that customers required, the test machine may
be used as the burning room, thermostat room and
thermostat-humidistat room if it is equipped with the
accurate control system of microcomputer for test room
have two kinds to be selected: one is the stainless steel and
the other one is the sheet baked by paintings.
Model description: H243-LPVST
H
Width
A=900mm B=1800mm C=2700mm D=3600mm E=4500mm F=5400mm
G=6300mm H=7200mm I=8100mm
24
Height
21=2100mm 24=2400mm 27=2700mm 36=3600mm
3
Deep
1=1800mm 2=2700mm 3=3600mm 4=4500mm 5=5400mm
6=6300mm 7=7200mm 8=8100mm
L
Temp. Range
B=rm. temp. +5°C~70°C H=0°C~70°C F=20°C~70°C
G=30°C~70°C L=40°C~70°C T60°C~70°C
P
G=set value of thermostat and humidistat, P=programmable thermostat and
humidistat, C=thermostat room
V
O=thermostat&humidistat control by set value, T=T-type programmable thermostat
and humidistat control, E=E-type Programmable thermostat and humidistat control
humidistat room, M=M-type programmable thermostat and humidistat control,
N=thermostat control by set value, V=V-type programmable control, P=P-type
programmable control, H=H-type programmable control
S
Outside material
, S=stainless steel, T=sheet coated with resin
T
Inside material
, S=stainless steel, T=sheet coated with resin
Construction
Thermostat and humidity
Thermostat room
Heat engine room
Temperature Range
H=0°C~70°C L=40°C~70°C F=20°C~70°C
T=60°C~70°C G=30°C~70°C
B=RT. +5°C~70°C
Control accuracy
±0.3°C ±3%RH
±0.3°C
Accuracy of distribution
±1°C ±5%RH
±1°C
Circulatory system
Convection by fan forced circulation Convection by fan forced circulation
with an additional hot air exhausting unit
Humidifying system
Vaporized from surface
Heating system
Heat dissipated by stainless steel
Dehumidifying system
Desiccated by frozen latent heat
Freezing system
Heat desiccated by way of air or water cooling with high effective freezing unit
Power source
AC220V/AC380V, 50/60Hz
Safety device
Leakage and overload protection device, compressor overload protection device,
power-off device for over-temperature and over humidity, water insufficient
protection device, over-temperature protection, device for humidifier and the
protection device for limit of temperature
H243-LPVST, The Walk-In Environmental Test Room
H243-LPVST
Walk-In Room
Humidity Controllable Range (at room temp. 20
°
C)
98% RH
100
90
80
70
60
50
40
30
20
10
0 10 20 30 40 50 60 70 80 90 100
TEMP
°
C
General Type
Low Humidity Type
Temperature & Humidity Range Table
Relatine Humidity %
1...,227,228,229,230,231,232,233,234,235,236 238,239,240,241,242,243,244,245,246,247,...738