MRC-2013-Catalog - page 217

217
E
Transilluminator, Dryers
ELECTROPHORESIS
Model
UV-26
UV Lamps
5-302nm lamp(8W) &
4x365 lamp(8W)
Intensity
Variable intensity 5% ~ 100%
Filter Size(mm)
L260xW210
UV Safety Cover Size(mm)
L320xW280
Dimensions(mm)
L340xW290xH85
Weight
5kg
Power
220V/50Hz
UV-26
UV safety cover
Filter Workface
Intensity knob
Power &
Wavelength Switch
GED-3121/GED-4535
GED-3121/GED-4535, Midi Gel Dryer
The microprocessor of Midi Gel Dryer controls temperature
from ambient +5°C to 90°C with timer from 0 to 999 mins
to heat gels from base plate and connect with a vacuum
pump to remove the moisture to dry gels homogeneously.
The surrounding groove design and the optimal sealing
silicon rubber with mask ensure the optimal sealing when
connecting with a vacuum pump.
Features:
Drying Area: 210x310mm (LxW), 350x450mm (LxW)
Microprocessor for temperature control
User temperature
calibration
Timer with alarm
2 digital LED display.
Model
GED-3121
GED-4535
Drying Area (mm)
L210xW310
L350xW450
Display
Two of 4 digital LED
Control
Digital Microprocessor controller
Temperature Control Range
ambient to 90°C
Temperature Increment
0.1°C
Temperature Uniformity at 37°C
±0.2°C
Temperature Calibration
Yes
Timer
0~999 mins
Operating Temperature
ambient to 40°C
Weight
approx. 9.0kg
approx. 15.0kg
Specifications:
UV-26, Dual-Wavelength UV
Transilluminator, 302nm/365nm
UV Transilluminators are designed for viewing
DNAIRNA in agarose gels stained with ethidium
bromide(EB) or other dyes.
Transparent hinged cover, with UV-safety cover to
protect users from exposure to hazardous UV
radiation, is mounted by slip-hinges for easily
adjusting to user’s visual angle.
UV-26 Dual-wavelength UV transilluminator provides
a uniform & intense source of 302nm & 365nm
ultraviolet lights to back-illuminate transparent
fluorescent materials.
Nucleic acids stained with fluorescent dyes such as
ethidium bromide and acridine orange can be
detected under UV illumination.
The filter surface is 210 x 260 mm. The light intensity
level can be set from 0%-100% continuously. Either
transilluminator can be converted from 302l365nm
to 254nm by replacing the UV lamps as described in
the Care and maintenance section.
1...,207,208,209,210,211,212,213,214,215,216 218,219,220,221,222,223,224,225,226,227,...738