Inlet LinerPackingTool
• Positionwool reproducibly every time.
• Accurate placement to a specific,measureddepth, up to100mm.
Recommended for inlet linerswith an ID
2mm.
Loosen the nut
on the side of
the tool and
adjust the
gauge to the
manufacturer’s
recommended
depth.
Eliminate
user
variation!
Place a
one-cen-
timeter
plug of
loosely
boundwool
at the top
of the inlet
liner.
Insert the liner
packing tool into
the liner until the
tool bottoms out.
Remove the tool.
Thewool is now
positioned cor-
rectly in the liner
and the liner is
ready for use.
Description
qty.
cat.#
price
Inlet Liner PackingTool
ea.
20339
$62
InjectionPortRepairTool
• Resurfaces critical inlet seal areas.
• ForAgilent split/splitless injectionports.
Removecontaminants,
achieveabetterseal!
Description
qty.
cat.#
price
Injection Port RepairTool
ea.
21393
$84
Replacement SandingDisks (5 fine& 5medium)
10-pk.
22689
$15
The critical seal at the base of
a split/splitless injector allows
a seal to form between the
injection port and the inlet
liner.This critical seal wears
over time andmay become
scratched or pitted, which com-
promises the sealing ability of
the injector. Use the Restek
injection port repair tool to
easily resurface the inlet seal
and remove contaminants; it
saves time andmoney by pre-
venting leaks.
aluminum oxide sanding disk (resurfaces critical
seal at base of injector)
injection port
bore cleaner
InjectorWrench forAgilent5890/6890/6850GCs
• Use to remove the septumnut andweldments duringGCmaintenance.
• High-quality stainless steel construction.
• Meets original equipment performance.
Use the smaller
end to remove
the septum nut.
Use the larger end to
tighten the split/split-
lessweldment nut.
Description
Similar to
Agilent part# qty.
cat.#
price
InjectorWrench for Agilent
5890/6890/6850GCs
19251-00100
ea.
22065
$28
Capillary InstallationGauge forAgilent5973MS
• Pre-seats ferrule onto column for consistent installations.
• Made fromhigh-quality stainless steel.
Description
Similar to
Agilent part# qty.
cat.#
price
Capillary InstallationGauge for Agilent
5973MS
G1099-20039
ea.
21894
$178
Install the nut and ferrule onto the
column, then insert the column through
the installation tool, exposing several
centimeters at the exit end. Tighten the
nut.
Score and remove the
exposed end of the
column, then loosen
the nut.
The ferrulewill be proper-
ly seated and should
remain in placewhen light
force is applied. Install
the column into the
GC/MS interface.
Capillary InstallationGauge
• Pre-seats graphite* ferrule onto column for consistent installations.
• Prevents crushed column ends.
• Made fromhigh-quality stainless steel.
*Recommended for usewith graphite ferrules.
Description
qty.
cat.#
price
Capillary InstallationGauge for Agilent-style fittings
(compact ferrules)
ea.
21034
$65
Capillary InstallationGauge for
1
/
16
" fittings
(
1
/
16
" ferrules)
ea.
21399
$65
Install the column nut and ferrule onto
the capillary column. Cut the column
end squarely. Slide the column into the
installation gauge to the recommend-
ed insertion distance. Finger-tighten
the column nut.
Tighten the assembly
withmoderate force to
ensure a properly seated
ferrule. Loosen the
assembly and remove the
column and column nut
with seated ferrule.
The ferrulewill be properly
seated in the column nut,
and should remain in place
when light force is applied.
If it slides loosely on the col-
umn, repeat procedure.
GCAccessoriesOrganizer forAgilent5890/6890andVarianGCs
• Ideal forkeepingGCaccessories and suppliesorganizedand easy to find.
• Built-in syringe and vial holders.
• Mounts on theGC for easy access.
• Includes allmountinghardware.
GC accessories
and supplies not
included.
Description
qty.
cat.#
price
GCAccessoriesOrganizer for Agilent 5890/6890 andVarianGCs ea.
22681
$140
800-356-1688
• 814-353-1300 •
Easilypre-seat ferrules for consistent installations!