Compound

Concentration

1. C7

61%

2. ethanol

10%

3. benzene

5%

4. 2-butanol (IS)

4%

5. toluene

20%

Column: D3606 TCEP Column Set-Up (Column 1: 5',

1

/

8

"

OD, 10% OV 101 on Chromosorb PAW 80/100, Column 2: 5',

1

/

8

"

OD, 20%

TCEP on Chromosorb PAW 80/100, and Column 3: 15',

1

/

8

"

OD, 15% Carbowax 1540 on Chromosorb W); Instrument: Agilent

6890; Sample: mixed standard prepared in C7; Inj.: 1 µL, direct; Inj. temp.: 200 °C; Carrier gas: helium, constant flow; Flow

rate: 26 mL/min.; Oven temp.: 135 °C, isothermal; Det.: TCD @ 200 °C

Poor resolution; ethanol

tails into benzene.

GC_PC01105

Figure 1

TCEP columns often fail to adequately resolve benzene from ethanol,

resulting in poor quantitative results

Compound

Concentration

1. C7

61%

2. ethanol

10%

3. benzene

5%

4. 2-butanol (IS)

4%

5. toluene

20%

Column: D3606 Application Column (2 column set, cat.# 83606) (Column 1: nonpolar Rtx

®

-1, 6' (1.8 m),

1

/

8

" OD, 2.0 mm ID and

Column 2: proprietary packing material, 16' (4.9 m),

1

/

8

" OD, 2.0 mm ID); Instrument: Wasson 3606/5599 combo 6890N GC;

Sample: mixed standard prepared in C7; Inj.: 1 µL, direct; Inj. temp.: 200 °C; Backflush time: 3 min.; Carrier gas: helium, con-

stant flow; Flow rate: 25 mL/min.; Oven temp.: 135 °C, isothermal ; Det.: TCD @ 200 °C

GC_PC01103

Baseline resolution of

ethanol and benzene!

Figure 2

Restek's new D3606 column set accurately and reliably separates benzene

from ethanol, improving quantitative accuracy.

2.5

5.0

7.5

12.5

15.0

17.5

Time (Min)

10.0

GC_PC01096

1

2

3

4

Compound

Concentration

1. ethanol

9.94 volume %

2. benzene

1.14 volume %

3. 2-butanol (IS) 4.01 volume %

4. toluene

9.93 volume %

Column: D3606 Application Column (2 column set, cat.# 83606-800) (Column 1: 6' (1.8 m),

1

/

8

" OD, 2.0 mm ID, nonpolar Rtx

®

-1

and Column 2: 16' (4.9 m),

1

/

8

" OD, 2.0 mm ID, proprietary packing material); Sample: gasoline; Inj.: 1 µL; Backflush: 3.0 min.;

Carrier gas: helium; Flow rate: 20.4 mL/min.; Oven temp.: 135 °C; Det.: TCD;

Courtesy of Joaquin Lubkowitz, Separation

Systems, Gulf Breeze, Florida

Figure 3

Ethanol and benzene are reliably resolved in commercial gasoline by

beta testers using the D3606 column set.

Using Restek gas filters is an easy way

to increase GC column lifetime. Our

triple gas filters protect your column

by removing hydrocarbons, oxygen,

and moisture.

TECH

TIP!

Increase column lifetime!

www.restek.com

13

Restek Super-Clean

Gas Filter Kit

• High-purity output ensures 99.9999% pure

gas (at max. flow of 2L/min.).

• “Quick connect” fittings for easy, leak-tight

cartridge changes.

• Glass inside to prevent diffusion; polycar-

bonate housing outside for safety.

• All traps measure 10

5

/

8

" x 1

3

/

4

" (27 x 4.4 cm).

• Each base plate unit measures 4" x 4" x 1

7

/

8

"

(10.2 x 10.2 x 4.8 cm).

Description

qty.

cat.#

Carrier Gas Cleaning Kit

Includes: mounting base plate,

1

/

8

" inlet/outlet fittings, and

oxygen/moisture/hydrocarbon

Triple Gas Filter

kit

22019

Lab Gas Issues?

Ensure Quality and

Maintain Production

with a Restek Solution.

Visit

www.restek.com/gas

Website :

www.chromtech.net.au

E-mail :

info@chromtech.net.au

TelNo : 03 9762 2034 . . . in AUSTRALIA