Table of Contents Table of Contents
Previous Page  96 / 298 Next Page
Information
Show Menu
Previous Page 96 / 298 Next Page
Page Background

• 2 •

www.restekcorp.com

Pinnacle II

New HPLC Columns, cont.

Column:

Pinnacle II C18

Catalog#:

9214565

Dimensions: 150x4.6mm

Particle size: 5µm

Pore size:

110Å

Conditions:

Mobile phase: A: 1% acetic acid in water

B: acetonitrile

Time (min.):

% B

0

50

2

50

9

80

12

80

13

50

Flow:

1.0mL/min.

Temp.: ambient

Det.:

UV @ 280nm

Peak List:

1. nordihydrocapsaicin

2. capsaicin

3. dihydrocapsaicin

Sample:

habanero nugget extract

Inj.:

10µL

Conc.:

4000µg/mL

Solvent: ethanol

0 1 2 3 4 5 6 7 min.

1

2

3

Figure 3

The Pinnacle II

C18 column achieves good resolution of three target capsaicinoids.

LC_0160

Nutraceutical and Food Applications

The Institute for Nutraceutical Advancement (INA)

has published a series of methods for the determina-

tion of active compounds in nutraceutical products.

Pinnacle II

HPLC columns exhibit excellent peak

shapes and reproducibility for natural product analy-

ses (Figure 2).

As society's taste in spicy foods grows, so does the

need to test and classify products and raw materials

for their heat levels. AOAC Method 995.03 specifies

the separation of the three target capsaicinoids

responsible for the heat in chili peppers. Using a gra-

dient elution program, an efficient separation of

these compounds is achieved with the Pinnacle II

C18 column (Figure 3).

Figure 2

The Pinnacle II

C18 column achieves excellent separation of hyperforin and adhyperforin

when analyzing a St. John’s Wort capsule.

Exclusive!

Restek offers the

Trident Direct

guard cartridge system

.

All of our Pinnacle

II phases are available

in guard columns. The Trident Direct

system features three levels of protection

against many of the contaminated samples

that analysts inject into a system.

Many New Phases Coming Soon!

Additional stationary phases will be

joining this family of products.

Base-Deactivated Columns Coming

Soon!

A line of columns that offer the structural

integrity of Type A silica, combined with

an improved treatment for base

deactivation. Excellent for analyzing

basic compounds.

Restek also offers a full line of HPLC

accessories—everything from

PEEK

fittings to syringes.

Peak List:

1. hyperforin

2. adhyperforin

Sample:

St. John's Wort capsule

Inj.:

20µL

Conc.:

4170µg/mL

Solvent: methanol

Column:

Pinnacle II C18

Catalog#:

9214565

Dimensions: 150x4.6mm

Particle size: 5µm

Pore size:

110Å

Conditions:

Mobile phase: acetonitrile:0.01N

phosphate buffer, pH 2.5

(15:85, v/v)

Flow:

1.2mL/min.

Temp.:

30°C, autosampler at 15°C

Det.:

UV @ 270nm

LC_0175