Capillary GC stationary phases for organochlorine pesticides (e.g. US EPA

Method 8081) must possess the selectivity needed to resolve target pesticides,

yet withstand the rigors of repeated injections of extracts containing harsh

residuals from the sample matrix. The US EPA method also requires a suitable

counterpart column for confirmation, to quantify potential pesticide “hits.”

An efficient way to meet these requirements is through dual column analysis,

which eliminates the need for a separate confirmation run, or GC/MS analy-

sis. A newly enhanced version of our Rtx®-CLPesticides2 column, with

dimensions of 30m x 0.32mm ID x 0.50µm (cat.# 11325) is now available for

this application. The Rtx®-CLPesticides2 column, coupled with an Rtx®-440

column of the same dimensions (cat.# 12939), can provide a complete sepa-

ration of the 20 most commonly analyzed organochlorine pesticides, listed in

Method 8081, in less than 12 minutes (Figure 1).

The unique selectivities of the Rtx®-CLPesticides2 column and the Rtx®-440

column enable the analysis to be run quickly, with good resolution and peak

shapes, as shown in Figure 1. Both columns produce similar run times, with a

set of elution order inversions and, because they share a common stationary

phase thickness, they should exhibit similar life expectancies. For accurate,

time-saving analyses of organochlorine pesticides, we highly recommend the

Rtx®-CLPesticides2 / Rtx®-440 column combination.

Rapid, Dual Column Analysis for Organochlorine Pesticides

12-Minute Analysis Using Rtx®-CLPesticides2 / Rtx®-440 Capillary GC Columns

By Jason Thomas, Environmental Innovations Chemist

• Analysis and confirmation with a single injection.

• Rapid analysis increases throughput.

• New, thicker Rtx®-CLPesticides2 phase increases column lifetime.

Rtx®-CLPesticides2 Column (fused silica)

ID

df (µm) temp. limits

length cat. #

0.32mm 0.50 -60 to 320/340°C 30-Meter 11325

Rtx®-440 Column (fused silica)

ID

df (µm) temp. limits

length cat. #

0.25mm 0.25 20°C to 320/340°C 30-Meter 12923

SeCure™“Y”Connector Kit

SeCure

“Y” connector body, 3 knurled nuts,

“Y” Universal Press-Tight

®

union, 3 ferrules.

Ferrules Fit Column ID qty.

cat.#

0.28/0.32mm

kit

20277

Organochlorine Pesticide Mix AB #2

20 components, listed in Figure 1, 8-80µg/mL in hexane:toluene

(1:1), 1mL/ampul

cat. # 32292

decachlorobiphenyl

2,4,5,6-tetrachloro-

m

-xylene

Pesticide Surrogate Mix

200µg/mL each in acetone, 1mL/ampul

cat. # 32000

Figure 1

Sub-12-minute dual column analysis of organochlorine pesticides,

using an Rtx®CLPesticides2/Rtx®-440 column pair.

GC_EV00777

1. 2,4,5,6-tetrachloro-

m

-xylene (surr.)

2.

α

-BHC

3.

γ

-BHC

4.

β

-BHC

5.

δ

-BHC

6. heptachlor

7. aldrin

8. heptachlor epoxide

9.

γ

-chlordane

10.

α

-chlordane

11. endosulfan I

12. 4,4'-DDE

13. dieldrin

14. endrin

15. 4,4'-DDD

16. endosulfan II

17. endrin aldehyde

18. 4,4'-DDT

19. endosulfan sulfate

20. methoxychlor

21. endrin ketone

22. decachlorobiphenyl (surr.)

GC_EV00776

Column:

Rtx

®

-CLPesticides2 30m, 0.32mm ID, 0.50µm (cat.# 11325) and Rtx

®

-440 30m, 0.32mm ID,

0.50µm (cat.# 12939) with 5m x 0.32mm ID Siltek

®

treated guard

tubing (cat.# 10027), connected using SeCure

“Y” Connector Kit (cat.# 20277) with

Universal “Y” Press-Tight

®

Connector

Sample:

Organochlorine Pesticide Mix AB #2, 8-80µg/mL each component in hexane/toluene (cat.# 32292),

Pesticide Surrogate Mix, 200µg/mL each component in acetone (cat.# 32000)

Inj.:

0.5µL splitless (hold 0.75 min.), 3.5mm Siltek

®

treated single gooseneck inlet liner (cat.# 20961)

Inj. temp.:

250°C

Carrier gas:

hydrogen, constant pressure

Linear velocity: 60cm/sec. @ 300°C

Oven temp.:

110°C (hold 0.5 min) to 230°C @ 25°C/min., to 330°C @ 15°C/min. (hold 1 min.)

Det.:

ECD @ 330°C

Instrument:

Shimadzu GC-2010 FID/ECD with AOC-20i autosampler

for

more

info

For many additional pesticides

reference mixes, please refer to

our catalog, or visit our website.