Table of Contents Table of Contents
Previous Page  291 / 298 Next Page
Information
Show Menu
Previous Page 291 / 298 Next Page
Page Background

17

800-356-1688 •

www.restek.com

2005 vol. 3

• Reliable thermal conductivity leak detector—

every analyst should have one.

• Compact, portable, ergonomic design—easy to

hold and operate.

• Sensitive—detects helium or hydrogen at 1x10

-4

cc/sec*.

• Fast results—responds to leaks in less than 2

seconds, zeros with the touch of a button.

• Built-in rechargeable battery—charging adaptor

included.

Caution: The Restek Electronic Leak Detector is NOT designed for determining leaks of combustible gases. A combustible gas

detector should be used for determining combustible gas leaks in possibly hazardous conditions.

*Sensitivity measured using helium.

In continuing our efforts to provide chromatographers with the best available columns, tools, and acces-

sories, we have enhanced our popular Restek Electronic Leak Detector. New features include internal bat-

tery charge capability, a low-battery indicator, a battery charge indicator light, yellow lights to signal a

nitrogen leak, a repositioned on/off switch, to eliminate accidentally powering on the unit, and a new

probe tip design that prevents debris from entering the unit. The new leak detector retains the microchip

technology that enables high sensitivity in a compact unit, the autozero feature that allows instantaneous

zeroing with the touch of a button, and the ergonomic design that puts all controls at your fingertips, for

maximum ease of use.

The new Restek Electronic Leak Detector is the affordable solution for detecting helium, hydrogen, or

nitrogen leaks in your GC system. Leaks can cause detector noise and baseline instability, waste carrier gas,

and shorten column lifetimes. The leak detector responds in less than 2 seconds to leaks of gases with ther-

mal conductivities different from air, indicating leaks with both an audible alarm and an LED readout.

The leak detector detects minute gas leaks that can go undetected by liquid leak detectors. And, remem-

ber—you should never use liquid leak detectors on a capillary system, because liquids drawn into the sys-

tem through the leaks will contaminate the system.

Description

qty.

cat.#

Leak Detector with 110Volt Battery Charger

ea.

22451

Leak Detector with 220Volt European Battery Charger

ea.

22451-EUR

Leak Detector with 220Volt UK Battery Charger

ea.

22451-UK

Leak Detector Accessory Kit

Leak Detector is easily

accessed when stored in

the mounting bracket.

The kit includes an adaptor fitting that fits over the probe assembly to detect very small leaks in hard-to-

reach locations, and a mounting bracket that can be affixed to the wall or GC.

Description

qty.

cat.#

Leak Detector Accessory Kit (adaptor fitting for probe, mounting bracket)

kit

22453

Enhanced Electronic Leak Detector

A Leak-Free System Stabilizes Baselines and Lengthens Column Life

By Donna Lidgett, GC Accessories Product Marketing Manager

new

for 2005

tech

tip

Avoid poor chromatography

caused by leaks—check for

leaks with the Restek Leak

Detector

Easy-to-clean

probe assembly.

Verify pinpoint leaks