American Society for Testing and Materials test
method D-3606-99 is focused on measuring ben-
zene and toluene in finished motor and aviation
gasolines: benzene can be determined from 0.1%
to 5% by volume and toluene from 2.0% to 20%
by volume. The method requires two columns
connected in series. Typically, column A is a 0.8
meter x 1/8 inch stainless steel packed column
containing a 10% loading of a nonpolar stationary
phase, such as Rtx
®
-1 or OV®-101. This column
separates sample components by boiling point.
After n-octane (C8) elutes, the column is back-
flushed to prevent heavier compounds from enter-
ing column B, the main analytical column. The
light compounds, C8 and below, pass into column
B, a 4.0 meter x 1/8 inch stainless steel packed col-
umn containing highly polar 1,2,3 tris(2-cya-
noethoxy) propane (TCEP). Here, the aromatic
compounds are separated from the non-aromat-
ics, and quantitative information is obtained.
For method D-3606-99, micropacked column
technology is an efficient, practical, time-saving
alternative to 1/8 inch packed columns.
Micropacked column A is a 0.8 meter x 1/16 inch
x 1mm ID Silcosteel
®
column packed with 10%
Rtx®-1 crosslinked on 60/80 Chromosorb
®
WAW.
Micropacked column B is a 4.0 meter x 1/16 inch x
1mm ID Silcosteel
®
column packed with 20%
TCEP on 80/100 Chromosorb
®
PAW. We installed
the columns in an Agilent 6890 GC capillary inlet,
configured in the on-column injection mode using
our Micropacked Column Adaptor Kit for On-
Column Injection (cat.# 22425). We used hydro-
gen as the carrier gas and, to attain the 9mL/min.
flow rate, we adjusted the column head pressure to
44psig at 145°C.
Figure 1 illustrates the analysis of a sample con-
taining 1% benzene, 2% toluene, and internal
standard methyl ethyl ketone (MEK), in
n
-pentane
(C5). The cycle time, just under 3 minutes, is
greatly reduced, relative to the 8 minute cycle
imposed by 1/8 inch packed columns and helium
carrier gas. The micropacked column / hydrogen
carrier gas combination reduces analysis time by
63%, nearly tripling sample throughput. If you are
performing method D-3606-99 analyses, and time
is important to you, we highly recommend this
micropacked column approach.
Rapidly Determine Benzene and Toluene in Gasolines
Micropacked GC Columns Reduce Analysis Time by 63%
By Barry Burger, Petroleum Applications Chemist
• 3-Minute Cycles for ASTM Method D-3606-99.
• Nearly triple sample throughput.
• Easy set-up, using Restek adaptor kit.
Figure 1
Benzene, toluene, and internal standard resolved in 3
minutes, using a micropacked column.
Micropacked Columns
1/16 inch micropacked columns containing 10% Rtx
®
-1 on 60/80
Chromosorb
®
W or 20% TCEP on 80/100 Chromosorb
®
PAW are prepared on
request. For details, please contact your Restek representative.
Micropacked Inlet Conversion Kits
Convert a capillary GC split/splitless inlet for use with
1
/
16
" OD micropacked
columns.
• For use with Agilent 5890 and 6890 GCs.
• Sample pathways deactivated for ultimate inertness.
*For use with packed column FIDs only.
Description
qty.
cat.#
Micropacked Column Adaptor Kit for On-Column Injection*
Complete kit with FID and injection port adaptors
Kit includes: Dual Vespel
®
Ring Inlet Seal, large bore; reducing nut, large
bore; FID adaptor, large bore;
1
/
4
" ferrule, Vespel
®
/graphite;
1
/
4
" nut,
stainless steel;
1
/
16
" ferrules, Vespel
®
/graphite (2); Siltek
®
-treated metal
liner installation guide;
1
/
16
" nuts, stainless steel (2)
kit
22425
1. non-aromatics
2. benzene
3. methyl ethyl ketone (int. std.)
4. toluene
GC_PC00775
Column:
A: 10% Rtx
®
1 on 60/80 Chromosorb
®
WAW, 0.8m x
1
/
16
inch x 1mm ID Silcosteel
®
tubing
B: 20% TCEP on 80/100 Chromosorb
®
PAW, 4.0m x
1
/
16
inch x 1mm ID Silcosteel
®
tubing
connected in series and using Micropacked Column On-Column Injection Kit (cat.# 22425)
Sample:
benzene, toluene, methyl ethyl ketone in pentane
Inj.:
1.0µL, on-column, 4mm ID splitless inlet liner (cat.# 20772)
Carrier gas:
hydrogen, constant flow
Flow rate:
9.0mL/min.
Inj. temp.:
200°C
Oven temp.:
145°C
Det.:
FID @ 200°C