Zero gas free calibration

With a unique design, you can avoid the need of providing zero

gas to the analyzer during calibration. By having a zero gas

generator in the analyzer, only span and sample gases are

required to operate the unit. The zero generator system ensures

stability at low ppb level. It also avoids presence of negative

readings seen in common trace nitrogen instrument when bad

zero calibration is performed.

An automatic system is managed by the software of the

analyzer. When choosing zero gas calibration, the zero gas

system is automatically selected to provide the zero reference

needed to calibrate it. The measurement can now be more

accurate than ever at low concentration without using an

external gas heated purifier.

This is also a cost effective configuration since savings are done

on the installation of the zero gas (tubing, valves, gas cylinder,

gas regulators, etc) and the need of an outside gas purifier.

PlasmaDetek

By using the PlasmaDetek

technology, the LD8000

is equipped with the state

of the art detector for

such analysis. Noise level,

accuracy and stability has

reached performance that

makes such trace nitrogen

analyzer the ideal tool for

any argon production plant,

laboratories or any other

needs.

Be sure to look at the LD8000 design report on our web site for

more details.

Crude argon analysis

The LD8000 can now measure trace nitrogen in a mixture

of argon and oxygen. The use of gas chromatograph is then

avoided and response time is faster by having reading

almost instantly.

A few ppm up to % N2 can be measured with this unique

device developed by LDetek. Now all CLAR (Crude Liquid Argon)

producers and users have the right tool to monitor the level of

this critical component.

The

state-of-the-art

trace nitrogen in argon,

helium or crude argon analyzer

LD

8000

LDetek has brought the measurement of trace

nitrogen to another level. The LD8000 is now an

instrument offering many possibilities that the

market was looking for a long time:

www.

ldetek

.com

Where

innovation

leads to

success

7