ANALYSIS OF PERMANENT GASES

IN DIFFERENT MATRIX USING

PLASMADETEK

This analyzer has been configured with two independent

channels, each one dedicated to one matrix. The system is based

on a GC Clarus 580 from Perkin Elmer equipped with electronic

pressure regulators. Two high quality 10 ports diaphragm valves

with internal purge ensure sample injection directly on-column

and matrix backflush.The precolumn is a semi-capillary Plot

Q 30m x 530um column. The analytical column is a Molsieve

5A 30m x 530um column with a by-pass flow to elute CO

2

. Two

detectors Plasmadetek with two internal detectors each are used

to quantify the components. The signal coming from each PED is

sent to the acquisition software Totalchrom.

Figure 1

Figure 2

Figure 3

Figure 4

Figure 5

A 5 ppm standard of each permanent gas has been used to

calibrate the system and determine LOQ for each component.

The chromatograms shown below illustrate the results we

can obtain from this configuration. The chromatogram in

figure 2 is obtained with the detector optimized for H

2

and CH

4

.

The chromatogram in figure 3 is obtained with the detector

optimized for N

2

, CO and CO

2

.

The use of Plasmadetek in this system provides excellent results

regarding sensibility and selectivity. The Plasmadetek is ideal

for this kind of analyses because it can support high volume of

sample injection and we can adjust Make Up gas flow to obtain

a suitable chromatogram. The installation of Plasmadetek on a

Clarus GC is easy and don’t need physical modification of the

GC. Its design is in full harmony with the GC whose all parts are

accessible for an easy modification (figure 5).

One of the main advantages of Plasmadetek is that it is

maintenance free, so that gives a robust system for long

analyses sequences in the laboratory or for on-line measuring

analyzers. The large range of concentrations (from ppb to %) and

the available selectivity of Plasmadetek make this detector ideal

to work in a lot of domains where GC is necessary.

Component Conc.

(ppm)

Peak

Height

Noise S/N LOD

(ppb)

S/N=3

LOQ

(ppb)

S/N=5

H

2

5

296 0.02 14800 1.0 1.7

CH

4

5

808 0.02 40400 0.4 0.6

CO

2

5

223 0.02 11150 1.3 2.2

N

2

5

793 0.02 39650 0.4 0.6

CO

5

110 0.02 5500 2.7 4.5

The goal of this application is to analyze permanent gases in different matrix used for electronic

industry. The investigation of these impurities is important for gas providers to ensure the high purity

of their products to their customers.