Reproducibility of Rtx-DHA-100

Columns

Following graphs show the typical performance of the

chronological production of 250 Rtx-DHA 100

columns

• Efficiency /Resolution

• Retention

• Inertness

• Bleed

Set by ASTM

Efficiency / Resolution ……Plate number

400000

450000

500000

550000

600000

650000

700000

1 13 25 37 49 61 73 85 97 109121133145157169181193205217229241

Method

specs

ASTMspec:C5 theoreticalplates :450.000 –550.000

ASTMspec:C5 theoreticalplates :450.000 –550.000

RestekRtx-DHA100 columns are coatedwith superior efficiency:

that’swhy theyalways separate better.. (datawillbemore accurate)

RestekRtx-DHA100 columns are coatedwith superiorefficiency:

that’swhy theyalways separatebetter.. (datawillbemoreaccurate)

0.4

0.42

0.44

0.46

0.48

0.5

0.52

0.54

1 13 25 37 49 61 73 85 97 109121133145157169181193205217229241

Method

specs

ASTMspec:C5 retention factor@35 ºC

: 0.45 -0.50

ASTM spec:C5 retention factor@ 35 ºC

: 0.45 -0.50

Retention …. Retention factor

Smalldeviation on retention factormeans similar retention times

column to column; Convenient to setup& calibrate

Smalldeviation on retention factormeans similar retention times

column to column; Convenient to setup& calibrate

0.4

0.42

0.44

0.46

0.48

0.5

0.52

0.54

1 13 25 37 49 61 73 85 97 109121133145157169181193205217229241

Method

specs

ASTMspec:C5 retention factor@35 ºC

: 0.45 -0.50

ASTM spec:C5 retention factor@ 35 ºC

: 0.45 -0.50

Retention …. Retention factor

Smalldeviation on retention factormeans similar retention times

column to column; Convenient to setup& calibrate

Smalldeviation on retention factormeans similar retention times

column to column; Convenient to setup& calibrate

2.5

3

3.5

4

4.5

5

5.5

6

6.5

1 13 25 37 49 61 73 85 97 109121133145157169181193205217229241

Resolution ……. R value

ASTMspec:Resolution

, t-butanol/2methylbutene2@35 ºC : 3.25 -5.25

ASTMspec:Resolution,

t-butanol/2methylbutene2@35 ºC : 3.25 -5.25

As theefficiencyand inertnessof theRestek column ismuchhigher the

resolution is alsomuch better then required..

As the efficiencyand inertness of theRestek column ismuchhigher the

resolution isalsomuchbetter then required..

Method

specs

ASTM D 6730 Requires a “tuned”

Column…

ASTM wants general purpose methods

Columns from different suppliers will show small

differences in selectivity..

By coupling a short piece of Rtx-5, every

column can be “tuned” to fulfill ASTM specs

on selectivity..

Tuning pre-column: Rtx-5, 2-5 meters

Length to be determined empirically