Important Quality Parameter for

DHA Analysis

• The column must not only elute hydrocarbons

• Also need to elute

alcohols

This is a big challenge and that’s why

Rtx-DHA 100 columns are preferred

1. ethanol

2. pentane

3. t-butanol

4. 2-methylbutene-2

Elution of Alcohols on the DHA-type

Columns ASTM Method D6730

Position of alcohols depend

on the activity of the surface..

Correct positions, Good deactivation

Column : 100m x 0.25mm PONA/DHA-type

Conditions : acc to ASTM 6730

Concentration: The SAME amount injected on every column

Rtx-DHA-100

Impact of a Tailing Alcohol Peak

Supplier 1 type DHA column

7.0

8.0

9.0

10.0

11.0

Time (min)

1

2

3

4

5

1. ethanol

2. C5

3. 2 methyl butene 2

4. t butanol

# 24160U

Oven: 5 ºC → 8.23 min., 22ºC/min → 48 ºC

Alcohol peaks are retained by column

activity / adsorption;

They elute at wrong position..

Risk: Misidentification and possible wrong data